Vacuum and Epics: Difference between revisions
Jump to navigation
Jump to search
Tag: Manual revert |
|||
(18 intermediate revisions by the same user not shown) | |||
Line 1: | Line 1: | ||
=Login= | |||
*To bring up epics, ask Annika or Martin for access to isacepics login. | *To bring up epics, ask Annika or Martin for access to isacepics login. | ||
=Navigation= | |||
*From the main menu, go to "vacuum2->HEBT-GP3" to open the ANASEN vacuum page (called HEBT-GP3 Vacuum) | *From the main menu, go to "vacuum2->HEBT-GP3" to open the ANASEN vacuum page (called HEBT-GP3 Vacuum) | ||
[[File:epicsmenu.png|600px|thumb|center|Epics menu]] | |||
[[File:hebtgp3.png|600px|thumb|center|HEPT-GP3 vacuum page]] | |||
*To view conditions of upstream beamline, click the "HEBT" button on the left of the page. | *To view conditions of upstream beamline, click the "HEBT" button on the left of the page. | ||
[[File:hebt3vacuum.png|thumb|center|600px|HEBT3 Vacuum]] | |||
=Pump down= | |||
[[File:HEBT_valves.jpg|500px|thumb|center|HEBT valves position.]] | |||
[[File:HEBT_valves2.jpg|500px|thumb|center|Manual valve - HEBT:MV19.]] | |||
===No window (fast pump down)=== | |||
*For fast pump down, make sure the manual valve is fully open | |||
*Close HEBT:VV19 (if open) | |||
*Start HEBT:RP19 and observe pressure on HEBT:CG19B | |||
*Once HEBT:CG19B is showing a reading below 50mTorr, open HEBT:PB19 | |||
* | |||
===With window (slow pump down)=== | |||
=Venting= | |||
==No window (slow venting procedure)== | |||
==With window (fast venting procedure)== |
Latest revision as of 01:54, 16 June 2022
Login
- To bring up epics, ask Annika or Martin for access to isacepics login.
- From the main menu, go to "vacuum2->HEBT-GP3" to open the ANASEN vacuum page (called HEBT-GP3 Vacuum)
- To view conditions of upstream beamline, click the "HEBT" button on the left of the page.
Pump down
No window (fast pump down)
- For fast pump down, make sure the manual valve is fully open
- Close HEBT:VV19 (if open)
- Start HEBT:RP19 and observe pressure on HEBT:CG19B
- Once HEBT:CG19B is showing a reading below 50mTorr, open HEBT:PB19